Arterial Stiffness

Evaluation of the Tensioday ambulatory blood pressure
monitor according to the protocols of the British
Hypertension Society and the Association for the
Advancement of Medical Instrumentation
Zso´ fia Ne´meth, Katalin Mo´cza´ r and Gyo¨rgy Dea´k
Background The validation of ambulatory blood
pressure monitoring devices is necessary to obtain
information on their accuracy. The objective of the
present study was to evaluate the accuracy of the
Tensioday oscillometric ambulatory blood pressure
monitor according to the protocols of the British
Hypertension Society and the Association for the
Advancement of Medical Instrumentation (AAMI).
Design We followed the phases recommended by the
British Hypertension Society protocol: before-use
calibration, in-use assessment, after-use calibration,
static device validation and report of the evaluation.
However, we expanded on the protocol to accommodate
features required by the AAMI.
Method The accuracy of calibration of three Tensioday
devices was tested before and after the in-use phase
when each of three devices was performing 10 24 h
sessions of ambulatory monitoring. As all three devices
passed these phases, the accuracy of blood pressure
measurement was tested in one arbitrarily selected
device on 85 subjects for systolic and 85 for diastolic
by suwataisya | 2014-05-28 17:22 | 動脈硬化